Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Re: Gasbench Nafion Traps
From:
Andrew Schauer <[log in to unmask]>
Date:
Mon, 16 Jun 2008 09:05:16 -0700
Content-Type:
text/plain; charset=us-ascii
MIME-Version:
1.0
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments:
text/plain (37 lines)
I believe Nafion traps are fairly robust and don't "break down". Certainly material could flow through them that would degrade them. If you suspect the ability for it to remove water, you might try lightly heating it while helium is flowing...you will typically get a fair bit of water off. Perhaps you had a wet helium cylinder? Are you detecting a leak? I have had bad luck with the base of the sample open split that seems to begin leaking after thousands of samples. 

andy


 Andrew Schauer
ISOLAB
Department of Earth and Space Sciences
University of Washington
Seattle, WA 98195


206.543.6327
[log in to unmask]


----- Original Message ----
From: Joe Lambert <[log in to unmask]>
To: [log in to unmask]
Sent: Monday, June 16, 2008 7:21:24 AM
Subject: [ISOGEOCHEM] Gasbench Nafion Traps

Hi everyone,

      I was hoping to get some advice regarding the two Nafion water traps
in Thermo's Gasbench.  Can anyone please explain the symptoms of a Nafion
trap that is no longer functional?  I'm assuming the trap directly after the
sample needle is typically the first to go.  Is there a way to test which
trap has the problem?  Finally, what is the expected life of Nafion traps? 
We've had about 7,000 samples through the machine with only a few hundred
being water samples.

Thanks in advance,

Joe Lambert
Alabama Stable Isotope Laboratory