Hi Magda & Diana--
After equilibration, we've taken to removing taking the tubes out (two
at a time) and tapping the sides of the cap to move the condensate to
one side, then shaking it down. There isn't much of a problem with the
runs afterwards. Alternatively, you may consider putting a blanket on
top of the gasbench while it's equilibrating to see if that cuts down on
condensation.
take care
gerry
Diana Bojic wrote:
> Dear All,
> We are new in mass spectrometry "business" but we already have
> problems. We are trying to do water equilibration on gas bench II and
> delta +XP. We have problem with evaporation of sample water in vials.
> Water is condensing on the septum and when the needle penetrate in
> septum small amount if water is entering the system. We have heard
> for building a cooler but we are interesting in otherer possibilities.
> Does anybody have some suggestions?
>
>
>
> Thanks
>
>
>
> Magda & Diana
>