Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Date:
Wed, 5 Jul 2006 20:49:17 -0400
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Re: ISODAT 2.0 & method
MIME-Version:
1.0
Content-Transfer-Encoding:
8bit
In-Reply-To:
<[log in to unmask]>
Content-Type:
text/plain; charset=ISO-8859-1; format=flowed
From:
Gerard Olack <[log in to unmask]>
Parts/Attachments:
text/plain (66 lines)
HI Joel--

You can even "duplicate" the Gasbench and give it a new name--and have 
the same problems.

The only work arounds I know of are to either go into ISL and "hardwire" 
it, e.g. always have autodilution kick in at 30V, or run another ISL 
program that saves the responses to the registry or file that 
"acquisition.isl" can read then read in.

By the way, if I recall correctly, the settings in the Method window, 
e.g. "autodilution", are saved in the Windows registry in a section you 
can't read from ISL. 

I hope someone else has a better answer.

good luck...

gerry


Joel Savarino wrote:

> Hello all,
>
> I have built a new configuration with all the needed valves, piston etc. I
> have addressed each valve and connected all control valves to the MS
> interface board. I have built my visual interface with my connecting 
> tube, actived all
> tools. I have tested and everything works fine, time events etc.
>
> Now, when I select my new config in Acquisition window and open a new 
> method file
> I don't get the gas bench options that normaly shows up with
> a gas bench method (e.i. autodilution, transfer time and extrascript) 
> on the first
> page of the method window. I have cross checked the gas bench config 
> with my new device,
> as the latter is very close to a gas bench config, and I don't see any 
> difference between the two
> devices (autosampler present, same options checked).
> Still when I create a new method, I don't get the same options in the 
> method window, such as the
> autodilution option for instance.
>
> did I forgot something in the configurator?
>
> Any help will be aprreciated.
>
> J
>
> *************************************************************************
> Joel Savarino
> Laboratoire de Glaciologie et Géophysique de l'Environnement/CNRS/UJF
> 54 rue Moliere, BP 96
> 38 402 St Martin d'Hères Cedex
> France
> tel: + 33 476 82 42 51
> fax: + 33 476 82 42 01
> Email: [log in to unmask]
> Web: http://www-lgge.obs.ujf-grenoble.fr 
> <http://www-lgge.obs.ujf-grenoble.fr/>
> **************************************************************************
>
>