Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Mime-Version:
1.0
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Vacuum problem solved
From:
Penny Higgins <[log in to unmask]>
Date:
Fri, 12 Jan 2007 16:10:46 -0500
Content-Type:
text/plain; charset="us-ascii"; format=flowed
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments:
text/plain (67 lines)
Dear isogeochemists,

Thanks all for your advice, both on and off line. I'd like to 
summarize the suggestions that were given to me, without naming names.

The problem: The rotary won't stop gurgling when trying to pump down 
after repair to one of the turbo pumps on a Delta Plus XP. Pumping 
aborts after 8 minutes because turbos can't reach 80% of their full speed.

The suggestions:
1) Check the ballast valve on the rotary pump (did it, was fine).
2) Degas the rotary pump (didn't do this - changed the rotary oil instead).
3) Check the oil wick and gasket on the turbo (didn't do this)
4) Let it go through the cycle of trying to pump down and timing out, 
then waiting for something to 'click' and try it again (the 'give it 
a chance, it'll figure it out' technique. Tried this one, no success.)
5) Check the screen between the inlet of the rotary pump and the rest 
of the vacuum plumbing. Clean it if oily (did this - it was an oily 
snarge. The thing still wouldn't pump down, however, after cleaning).
6) Check that the Penning Gauge is working properly (didn't try, but 
was on my list).
7) Check the seals on the down-to-air valve (took it out and reinstalled it).
8) Check that the down-to-air valve is actually functioning properly 
(wasn't sure how to do this - but the red light was on, indicating it 
was getting power and it did hiss whenever the pumps went off again 
after pumping for a while.)
9) Check the lights on the control module of the turbo pump - maybe 
the fault is there (All the happy lights were on and the not-happy 
lights were off).
10) Give the turbo a gentle whack (the 'teach it who's boss' 
technique? Was saving that for last.)
11) Make sure that the turbo is seated properly with the knive-edge 
gasket (did this with the source which I had had to pull out and 
twice with the turbo with no success. Tried once more with a 
brand-spankin-new seal on the turbo, and guess what? It worked!)

So in the end, it was my stupidity that caused it to fail. <<hanging 
head in shame>> Happily, though, I should be able to start running 
analyses again next week. Whew!!

Thanks to all who gave me suggestions! I hope this summary is helpful 
to others who may later have similar troubles.

~Penny


*******************************************************************
                       Dr. Pennilyn Higgins
                        Research Associate

                             "SIREAL"
   Stable Isotope Ratios in the Environment Analytical Laboratory

         Department of Earth and Environmental Sciences
         University of Rochester
         227 Hutchison Hall
         Rochester, NY 14627

                    [log in to unmask]

Office: 209b Hutchison Hall               Lab: 209 Hutchison Hall
Voice : (585) 275-0601              Outer lab: (585) 273-1405
FAX   : (585) 244-5689              Inner lab: (585) 273-1397

           http://www.earth.rochester.edu/SIREAL/index.html
*******************************************************************

ATOM RSS1 RSS2