Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Mime-Version:
1.0
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
HD by GC-IRMS
From:
Paul Eby <[log in to unmask]>
Date:
Wed, 17 Nov 2004 10:01:53 -0800
Content-Type:
multipart/mixed; x-avg-checked=avg-ok-7917AB8; boundary="=======4BFF4787======="
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments:
text/plain (1366 bytes)
List members,

Our lab has been working on hydrogen isotope measurements by GC.
Specifically, we have been measuring hydrocarbons anywhere from methane up
to C10, on a Delta Plus XL.

The problem is that extreme time shifts can be found between the mass 2 and
mass 3 signals: upwards of 2 seconds. This is using the same columns and
injection methods that we would for d13C, for which the time shift is
almost negligible.

I've been informed by Finnigan though that Isodat isn't capable of handling
time shifts that large. In order to stop the software from confusing
adjacent peaks as a time shift, they have imposed an artificial limit on
the time shift - it can only be as large as the integration time (default
is 0.25 seconds).

Finnigan is apparently working on a fix for this, but in the meantime I am
looking for other solutions. Increasing the integration time, for instance,
has limited usefulness as it also changes the integration and at 2 seconds
misses most of the peak.

Have others experienced this large time shift? Is it simply related to
chromatographic efficiency, or are there other factors that are involved?
Why do these large shifts appear when measuring hydrogen but not carbon?