Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Date:
Thu, 9 Jun 2005 11:42:58 -0600
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Re: laser cell design for carbonates
MIME-Version:
1.0
Content-Transfer-Encoding:
8bit
In-Reply-To:
<[log in to unmask]>
Content-Type:
text/plain; charset="ISO-8859-1"; format="flowed"
From:
Zachary D Sharp <[log in to unmask]>
Parts/Attachments:
text/plain (49 lines)
Dear Kim,

We have a graveyard of sample chambers of bewildering
dimesions.  The bottom line is that if you have a large
chamber, it takes more time to sweep all of the CO2 gas
out of the chamber into the line.  Also, if you have
larger samples, there is a much longer time to degas them.
 All in all, I think the solution is a small chamber.
 Take your large sample and break it up into smaller
pieces, loading one at a time.  With a small chamber, the
flushing time after exposure to air is so short, that you
should be able to change samples and be up-and-running
again in just a few minutes.

Sincerely,

Zach

On Thu, 9 Jun 2005 12:22:45 +1000
  Kim Baublys <[log in to unmask]> wrote:
> Isogeochemists,
>
> We are currently running a laser carbonate system
>(CF-IRMS with a 75watt
> CO2 Merchantek laser ).  We need to redesign the laser
>cell to be able to
> handle larger sample sizes.  We would like any
>information/advice on the
> properties and costs and people's experience with
>different cell windows
> and cell designs.  I've searched the archives and laser
>ablation of
> carbonates was last mentioned way back in 1995.
>
>
> Thanks
> Kim
>
>
> Kim Baublys
> Stable Isotope Laboratories
> Department Of Earth Sciences
> University of Queensland 4072
> Australia
>
> Ph:   +61 7 3365 1131
>Fax: +61 7 3365 1277
> http://www.uq.edu.au/geochem/