Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Show All Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Subject:
power needs for Delta V and peripherals
From:
"Lora L. Wingate" <[log in to unmask]>
Reply To:
Stable Isotope Geochemistry <[log in to unmask]>
Date:
Thu, 3 Apr 2008 16:54:43 -0400
Content-Type:
text/plain
Parts/Attachments:
text/plain (21 lines)
Hi all,
We are in the process of determining whether our existing 30kVA transformer 
will support our presently-installed instrumentation in addition to units 
that are slated to be installed soon.  Has anyone measured the actual amps 
pulled by a Delta V/Conflo/GasBench/TCEA combo?  Based on the 
manual-provided kW rating for the units, we calculate the amps the be 
(1)  Delta V baking, all peripherals powered up to be ~37A  (2) Power to 
Delta V without baking and peripherals powered up to be ~24A.   Does this 
sound ballpark for either the operational or baking state?
Thanks in advance.



Lora L. Wingate
University of Michigan
Department of Geological Sciences
Stable Isotope Laboratory
1100 North University
1013 C.C. Little Building
Ann Arbor, MI 48109-1005 