Conical vials for MultiPrep
From:
"Stephen S. Howe" <[log in to unmask]>
Reply To:
Stable Isotope Geochemistry <[log in to unmask]>
Date:
Wed, 22 Jun 2005 14:45:03 -0400
Content-Type:
text/plain
Parts/Attachments:
text/plain (34 lines)
Clemente and others:

Wheaton makes a conical-shaped vial (they call them V-vials) that has
a volume of 1.0 mL, an O.D. of 13 mm, and uses a 13-425 screw cap
with a hole in the center that can be lined with silicone rubber
septa and Kel-F liners.  Wheaton's part number is 986284 and can be
found through Fisher, VWR, etc., or probably directly through
Wheaton.  You should make sure that the vials will fit in the holes
of the heated rack of the MultiPrep or MultiFlow attached to the
Gilson 222 XL autosampler before you buy a big batch of them.  The
heated rack that was installed with our MultiPrep in 1996 had holes
that could accomodate Wheaton vials that were actually being made
with an O.D. very slightly smaller than 13 mm at that time.  Shortly
thereafter Wheaton's vials "returned" to their nominal O.D. of 13 mm
and no longer fit in our rack.  Fortunately, we purchased quite a few
of the smaller-diameter vials but they are starting to wear out.  The
alternative would be to have the holes of our heated rack bored out
very slightly, and eventually, we may have to resort to this remedy.

Hope this helps!

Steve


--
---------------------------------------------------------------------------
Stephen S. Howe                         Office:  (518) 442-5053
Dept. of Earth & Atmos. Sciences        Stable Isotope Lab:  (518) 442-4471
Earth Science 352B                      FAX:  (518) 442-5825
University at Albany                    E-mail:  [log in to unmask]
1400 Washington Avenue        http://www.albany.edu/geosciences/sshowe.html
Albany, NY  12222-0001      http://www.albany.edu/geosciences/sirmslab.html
---------------------------------------------------------------------------