Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Mime-Version:
1.0
Content-Type:
text/plain; charset=us-ascii
Date:
Thu, 14 Aug 1997 09:20:20 -0700
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Re: Oxygen isotope exchange between reference gas CO2 and a contaminant?
From:
Mahendra Pal Verma <[log in to unmask]>
Content-Transfer-Encoding:
7bit
Organization:
Inst. de Inv. Electricas
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments:
text/plain (21 lines)
Dear Paul Dennis,

We never faced such type of problem. But few things could be done to know the
causes for it.

1. Try to run same gas as standard as well as sample.

2. Adjust the pressure on the capillaries (SA and ST sides), so that the flow
is similar in both sides under similar conditions.

3. There could be some problem with filament. It could be replaced.

4. The last possibility is to check electronics, especially of Faraday cups
amplifiers. But it is really hard  to pin point about electronics faults.


Good Luck