Dear Paul,
about two years ago I had a similar "drift" problem with our dual
inlet Finnigan MAT251. You did'nt specify the type opf your mass-
spec, but in our case it helped tremendously to take the capillaries
(on both sides) out of the squeezers, widen the squeezed point
carefully with a pair of pliers (a little, just that bit!) and glow them
red hot with a gas flame. Whatever contaminant there was, it
disappeared while constantly pumping. The zero-enrichment test
came back to what it used to be.
If in fact your bellows are the problem, the spectromat people in
Bremen/Germany may have some good and affordable ideas how
to fix the problem. Contact Hansi Wust <[log in to unmask]>
regards to all,
Uwe Horstmann