Dear all,
We recently encountered the same problem with our Gasbench on the deltaplus
with IsodatNT software. Due to the impossibility to peakcenter, our
sequence will stop. Shutting down the Isodat software by double-clicking
the tray-icon and restarting Isodat will help. Otherwise rebooting is also
a possibility. We've been using our systems since three months now and this
problem popped up only the last week or so.
Our GC-IRM (combustion interface), running on the deltaplusXL however
doesn't give this problem and seems to run without any problems since the
beginning.
IsodatNT is an improvement, but if these bugs can be solved, it would make
things even better still.
Michiel
__________________________________________
Michiel Kienhuis
Dept. Marine Biogeochemistry and Toxicology (MBT)
Royal Netherlands Institute for Sea Research
P.O. Box 59, 1790 AB Den Burg, The Netherlands
Internet http://www.nioz.nl
Email: [log in to unmask]
___________________________________________