Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Content-Transfer-Encoding:
7bit
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Re: Vacuum problem
From:
Franz Buegger <[log in to unmask]>
Date:
Thu, 25 Mar 2004 11:08:40 +0100
Content-Type:
text/plain; charset=us-ascii
MIME-Version:
1.0
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments:
text/plain (47 lines)
Dear Ricardo,

I've similar experience as Li and Howard.
At my delta plus the pirani gauge was mounted in such a way that the gauge
pointed to the bottom, so the oil condensate accumulates there and destroyed
it over the time. If it is the same way at your delta plus my tip is to add
a angle before changing the pirane gauge -and I fear you have to- in such a
way that the pirani gauge points to the top to avoid coming in oil from the
rotary pump.

Franz

Ricardo De Pol Holz schrieb:

> Dear members,
>
> I have just changed my delta plus filament and reinstalled the Source. I
> also renew the seal gasket to avoid problems. When I started the pumps
> again, my Fore vacuum went only to 4.5e-1 mbar but that was enough to
> start the turbopumps. After 2 hours the fore vacuum lowered to 8.5e-1
> mbar but the high vacuum was in the good direction (2.5e-7 mbar). All
> leds at the front panel are "green" (on).
> Since I didn't manipulate any of the FV conections, I only opened the
> source I don't have any idea what could be happening...or perhaps it is
> normal after a source opening and I should only wait. It worries me a
> little because the rotary pump has the typical bubbling sound as if its
> recently started (but it usually desappeared after a little while before
> this problem).
>
> Plase...any hints are most welcome.
>
> Thanks a lot.
>
> Ricardo De Pol-Holz
> Universidad de Concepcion

--
Franz Buegger
Institute for Soil Ecology
GSF - National Research Center for Environment and Health
Ingolstaedter Landstr. 1
D-85764 Neuherberg/Germany

Tel.: +49-89-3187-3403
Fax:  +49-89-3187-3376
Email: [log in to unmask]