Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Show All Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Subject:
Re: Delta plus losing signal in only one cup foronly one run
From:
Darren Sleep <[log in to unmask]>
Reply To:
Stable Isotope Geochemistry <[log in to unmask]>
Date:
Fri, 1 Mar 2002 15:23:59 +0000
Content-Type:
text/plain
Parts/Attachments:
text/plain (22 lines)
I'm sure this is a software problem as you suggest but some years ago
on our Tracermass we had a similar problem with Minor 1 cutting out
unpredictably. This turned out to be caused by a screw which had
vibrated loose within the collector housing and was shorting the
middle collector to the housing. Due to the housing being circular the
screw rolled around and sometimes rolled away from the collector
resulting in the beam reappearing - this rather depended on how heavy
the operator was and where they stood on the lab floor!





Darren Sleep  GIBiol
Stable Isotope Facility
CEH Merlewood
Grange-Over-Sands
Cumbria LA11 6JU
UK
tel +44 (0) 15395 32264
fax +44 (0) 15395 34705