Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show HTML Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Content-type:
multipart/alternative; boundary="Boundary_(ID_eC7ki2zk/4thxjzIoT2W6A)"
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Re: Gilson autosampler problem
From:
Li Huang <[log in to unmask]>
Date:
Mon, 9 Jul 2007 15:39:55 -0400
In-Reply-To:
<[log in to unmask]>
MIME-version:
1.0
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments:
text/plain (1744 bytes) , text/html (3110 bytes)
Dear all,
 
A few weeks ago I had posted a message about our problem in hoping to get some tips from the community. The amount of help I got was overwhelming and I appreciate all that time you had spent on it. 
 
The good news is the problem had been solved. It turned out, like some of you suggested, the power supply to the sampler was about to go. It works fine after the replacement. 
 
The replacement was done by Mandell  in their company (not the greatest experience, but purpose was served.) and costed us Can. $ 754.
 
Thanks to you all for the great help and I appreciate it.
 
li
 
 
 
 
 
----- Original Message -----
From: Li Huang <[log in to unmask]>
Date: Thursday, May 24, 2007 3:48 pm
Subject: [ISOGEOCHEM] Gilson autosampler problem
To: [log in to unmask]

> Hi,
>  
> We have had Multiprep for a few years and are experiencing some 
> problem with Gilson autosampler. Here's what we see. In the 
> middle of the autorun, every time the arm reaches row 3, 4 or 5 
> ( our sampler has 5 row and 12 column.)  Gilson returns to 
> zero position right after the needle pulls out of the vial and 
> will not move to the next position without manually initializing 
> the Gilson. 
>  
> It is difficult for us to run autorun  with the problem 
> like this. It looks to me that there is problem on that section 
> of the belt ( or track the sampler arm is travel on.) but don't 
> know how to check it. I wonder if any body had seen this before 
> and how they got this problem solved.
> 
> Thanks,
> 
> li
>  
>  
>  
>  
> 
>  Li Huang
> 
> Laboratory for Stable Isotope Science
> University of Western Ontario
> London, Ontario
> Canada
> Phone: 1-519-661-3881
> 
> 

 Li Huang

Laboratory for Stable Isotope Science
University of Western Ontario
London, Ontario
Canada
Phone: 1-519-661-3881