| Mime-Version: |
1.0 |
| Sender: |
|
| Subject: |
|
| From: |
|
| Date: |
Tue, 3 May 2005 11:21:28 +0200 |
| Content-Transfer-Encoding: |
8bit |
| Content-Type: |
text/plain; charset="iso-8859-1" |
| Reply-To: |
|
| Parts/Attachments: |
|
|
Dear all,
using a GC-MS, we are re-oxidising the tube when copper is nearly
completely black (we check this regularly) by heating the tube from 650 to
750 °C and connecting to O2/N2 gas ("Protec") for 10-15 minutes, then
cooling down to 650°C (about 30').
This is enough for next about 200 samples (two days, depending on
re-aging). If we have to re-oxidise it every day, we change it completely.
One tube is working in average for about 1000 analysis.
We did not try to re-oxidise it for 12 hours; maybe we would not need to
re-oxidise it all two days. Good idea.
Greetings,
Deborah
********************
Deborah Rupprecht
Institute of Plant Nutrition
Karlrobert-Kreiten-Str. 13
53115 Bonn, Germany
Phone (+049) 0228-73-3640, -1631
********************
|
|
|