Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show Text Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Mime-Version:
1.0
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Trap and Box current
From:
Shapna Mazumder <[log in to unmask]>
Date:
Thu, 2 Oct 2008 21:00:18 -0400
Content-Type:
text/plain; charset="utf-8"
Content-Transfer-Encoding:
quoted-printable
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments:
text/plain (13 lines)
Hello,

Just installed the source of Delta V after cleaning it and putting in a new 
filament.  After vacuum reached 5e-07 I turned on the source and am getting 
a 1.5 mA current for both Trap and Box.  I cleaned our Delta Plus source 
previous to this one and everything was fine, I am used to seeing 
either .7/0.78 when everything is working well or 1.5 and 0 when its a bad 
filament etc.  What is happening, any help is greatly appreciated.  Thanks.

Best Regards,
Shapna Mazumder