Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Proportional Font
Show Text Part by Default
Show All Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Subject:
Re: AS 800 Autosampler
From:
Michael Elefante <[log in to unmask]>
Reply To:
Stable Isotope Geochemistry <[log in to unmask]>
Date:
Mon, 25 Apr 2005 11:58:32 -0400
Content-Type:
text/plain
Parts/Attachments:
text/plain (13 lines)
Hilary,

I am new to this discussion group, but I have 15 years of experience with
GC and GC MS systems.  I also have worked on the AS800 autosampler since it
was introduced.  As far as memory effects from the autosampler go, it is
really sample dependent. What I have always done is change the amout of
washes to see how many were needed, if that doesn't work, you may need to
use more than one solvent, or a different solvent.  If you are still have
trouble feel free to contact me.

Mike Elefante
Elf Analytical Services, Inc.

ATOM RSS1 RSS2