Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Forum View

Use Monospaced Font
Show HTML Part by Default
Condense Mail Headers

Message: [<< First] [< Prev] [Next >] [Last >>]
Topic: [<< First] [< Prev] [Next >] [Last >>]
Author: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Re: problem on EA auto-sampler
From:
Michael Elefante <[log in to unmask]>
Date:
Fri, 27 May 2005 16:24:26 EDT
Content-Type:
multipart/alternative; boundary="part1_8e.28392f85.2fc8dbfa_boundary"
MIME-Version:
1.0
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments:
text/plain (784 bytes) , text/html (954 bytes)
I have been working on the Carlo Erba\Fisons\Thermo EA systems for 15 years
and I am positive I can fix your problem. The first thing to check is the
indexing arm on the left side of sampler. There is a bracket around it and if it is
too far foward the indexing arm might not click into the next tooth on the
gear consistently. The way you describe the problem if it is always the second
sample of a run and always position two on the tray you might have a bad tooth
on the gear. If the problem is always the second sample but not always
position two on the tray you may just need to clean the Teflon slide. I service all
Carlo Erba\Fisons\Thermo EA systems and have any parts you may need. Contact me
with any questions or comments you may have.

Mike Elefante
Elf Analytical Services
732-922-8400