Sender:
Stable Isotope Geochemistry <[log in to unmask]>
Date:
Mon, 3 Aug 2009 16:51:13 -0700
Reply-To:
Stable Isotope Geochemistry <[log in to unmask]>
Subject:
Re: using dual inlet during conflow
MIME-Version:
1.0
Content-Transfer-Encoding:
quoted-printable
In-Reply-To:
A<[log in to unmask]>
Content-Type:
text/plain; charset="US-ASCII"
From:
"Harlow, Benjamin" <[log in to unmask]>
Parts/Attachments:
text/plain (55 lines)
Hi Greg,

You need to open configurator and add the Dual inlet to your conflo
configuration.  I can't open it right now but I think you still add it
to "source". 

You will have to make new methods to do anything once you change this.
It's a good idea to run version handler before and after the
modifications so you can restore.  With the new changes you should see a
lot more options in your methods screen.

If you are interested I also have a script to make the bellows adjust to
a set pressure/voltage before every sample.

Regards,

Ben

 
Benjamin Harlow
Manager, Stable Isotope Core Laboratory
Washington State University
School of Biological Sciences
G-81 Eastlick
Pullman, WA  99164-4236
 
Office:  509-335-6161
Lab:     509-335-6154
Fax:     509-335-3184
 
Laboratory for Biotechnology and Bioanalysis (LBB2) Stable Isotope Core
http://www.isotopes.wsu.edu
 
 

-----Original Message-----
From: Stable Isotope Geochemistry [mailto:[log in to unmask]] On
Behalf Of greg michalski
Sent: Monday, August 03, 2009 4:38 PM
To: [log in to unmask]
Subject: [ISOGEOCHEM] using dual inlet during conflow

>Any ideas on how to get Isodat to allow operation of the dual inlet 
>(or change over block) while running a conflow aquisition?  When 
>running a sequence a not compatible with current configuration error 
>pops up. i think this some conflict between devices in the "source" 
>(dual inlet) and "capillary" (conflo).

Greg Michalski, Assistant Professor
http://www.purdue.edu/eas/michalski/
Department of Earth and Atmospheric Sciences
Purdue University
550 Stadium Mall Dr.
Lafayette IN.47907