Skip Navigational Links
LISTSERV email list manager
LISTSERV - LIST.UVM.EDU
LISTSERV Menu
Log In
Log In
LISTSERV 17.5 Help - ISOGEOCHEM Archives
LISTSERV Archives
LISTSERV Archives
Search Archives
Search Archives
Register
Register
Log In
Log In

ISOGEOCHEM Archives

Stable Isotope Geochemistry

ISOGEOCHEM@LIST.UVM.EDU

Menu
LISTSERV Archives LISTSERV Archives
ISOGEOCHEM Home ISOGEOCHEM Home

Log In Log In
Register Register

Subscribe or Unsubscribe Subscribe or Unsubscribe

Search Archives Search Archives
Options: Use Classic View

Use Monospaced Font
Show HTML Part by Default
Condense Mail Headers

Topic: [<< First] [< Prev] [Next >] [Last >>]

Print Reply
Mime-Version: 1.0 (Apple Message framework v752.2)
Content-Type: text/plain; charset=US-ASCII; delsp=yes; format=flowed
Date: Wed, 29 Nov 2006 12:29:53 -0500
Reply-To: Stable Isotope Geochemistry <[log in to unmask]>
Subject: Re: Gasbench acid pump again
From: Marilyn Fogel <[log in to unmask]>
In-Reply-To: <[log in to unmask]>
Content-Transfer-Encoding: 7bit
Sender: Stable Isotope Geochemistry <[log in to unmask]>
Parts/Attachments: text/plain (46 lines)
We own one of those acid pumps. With a new Gas Bench, colleagues  
strongly advised us to go with the human approach--a needle, an arm,  
and a syringe. I think it works pretty well and comes under the  
category of "Good, Fast, and Cheap".

Marilyn Fogel


On Nov 29, 2006, at 3:34 AM, Geldern, Robert van wrote:

> Greetings,
> Finnigan changed the design of the acid needle in the Gasbench.
>
> The new needle is thinner and very flexible.
> However, the drops are smaller than using the old, thick needle.
> Due to the smaller drops it seems that more drops are necessary for  
> the
> complete reaction of the carboante sample. Dropping of 5 to 8 drops  
> does
> not work, since the acid volume transfererd into the vial is to small
> and the sample is not dissolved at all or only in parts.
>
> So, we incereased the number of drops up to 15. Consequently, the
> dropping times increases dramatically. This causes another problem:
> Due to the long dropping time the CO2 in the vial to be measured (with
> sample needle in) is already diluted by the helium flow into the vial.
> So the CO2 does not last for 10 sample peaks.
>
> I incereased the volume of the acid pump to maxium (450mikroliter per
> stroke). It still takes 20 strokes per drop!! So for 15 drops you will
> need 300 strokes...
>
> Any suggestions?
>
>
> Robert
>
>
> -- 
> Dr. Robert van Geldern
> Leibniz Institute for Applied Geosciences (GGA)
> Section 3: Geochronology and Isotope Hydrology
> Stilleweg 2, 30655 Hannover, Germany
> phone: ++49-511-643-2313
> mailto:[log in to unmask]