Hello Everyone, I have a problem with my mass spectrometer - EA system. When we run empty tin cups we are getting a higher than normal background for beam 44 (6 to 9 E-10). I've tried different batches, sizes and types of tin cups and still the same problem. Interestingly, when we run real blanks (i.e., no tin cup is dropped, but everything else remains the same) we don't see the high background. This leads me to believe that the tin cups are contaminated. Has anyone else noticed a problem with the tin cups supplied by Elemental Microanalysis Limited beginning in December 2002? Can anyone think of what else may be causing our high background if it is not the cups? I have changed the gases (O and H) and the He purifier, heated the GC column to give it a cleaning, changed the combustion column using both pre-packed columns and columns we packed but still the problem persists. I've checked for leaks on both the EA and mass spec and even cleaned the source. Any help would be greatly appreciated. Jennifer ------------------------------------------------------ Jennifer Mckay Agente de Recherche / Research Associate GEOTOP - Universite du Quebec a Montreal C.P. 8888, succursale Centre-Ville Montreal, Quebec, Canada H3C 3P8 Telephone bureau/office: 514-987-3000 (poste/ext 2413) Telephone lab: 514-987-3000 (poste/ext 7028) Telephone GEOTOP: 514-987-4080 Telecopieur/Fax: 514-987-3635 E-mail: [log in to unmask] or if that fails try [log in to unmask] ------------------------------------------------------