Folks, I am trying to react carbonates using a GV Multiflow (Basically same Idea as GasBench as far as I can tell in that the H3PO4 reaction occurs in a helium environment and not a vacuum. So both camps are free to answer), and I noticed that after the H3PO4 was dispensed there was a tiny bubble at the bottom of my sample vial. This is obviously not a good thing. Has anyone experienced similar issues with continuous flow carbonate analysis? Is this one advantage the running with a vacuum system such as the GV multiprep? I am using 1ml Wheaton vials that have a cone-shaped bottom and the H3PO4 has a sg. of 1.94. What bottom-shaped vials are people using with the GasBench or Multiflow? Maybe thats the problem. Reaction temp=90°C Thanks in advance, Toti. Toti Larson Ph.D. Los Alamos National Laboratory Earth and Environmental Sciences 505.667.9894 (lab) Los Alamos, New Mexico 87545 550.667.8006 (office)