Hi All, We have a Carlo Erba NA1500 elemental analyser (attached to an Isoprime) and the auto sampler just stopped during an analytical run and will not operate even on manual. By switching the compressed air lines over I can get the auto-sampler to close, and then switch lines again and get it to open, which shows that the auto sampler is working OK. This led me to believe that the solenoid valve operating the auto sampler was faulty. I have tested all the coils and they have the same resistance (150ohms), so I am assuming that the solenoid valve is faulty. Either that or it is not getting the command to operate. Does this sound like a reasonable assumption? Has anyone else had this problem before? Is there anyway to test the solenoid valve without running the furnaces up to temperature? Any advice welcome. thanks kim