Hi All,
Does anyone know how to close all the valves in the changeover block in a Finnigan 252 using Isodat 2.0?  I need them all closed for maintenance.   Seems like it should be do-able but they appear to have made the software completely dummy proof.
If there is no software solution, I figure I can just remove air pressure to the valves manually.  My assumption is that they are normally closed valves.  Can anyone confirm this?

Thanks, Jason

Jason Curtis, Ph.D.
Stable Isotope Mass Spec Lab manager
Senior Associate-In Geochemistry
Department of Geological Sciences
241 Williamson Hall
University of Florida
Gainesville, FL 32611
[log in to unmask], office 352-392-2296
lab 352-392-3344, fax 352-392-9294