Jason,
You can close each changeover valve to the waste turbo by
holding down the keyboard control key followed by clicking on the
valve.
Andy
Hi All,
Does anyone know
how to close all the valves in the changeover block in a Finnigan 252 using
Isodat 2.0? I need them all closed for maintenance. Seems like it
should be do-able but they appear to have made the software completely dummy
proof.
If there is no
software solution, I figure I can just remove air pressure to the valves
manually. My assumption is that they are normally closed valves. Can
anyone confirm this?
Thanks, Jason