Tim Heaton wrote: > Paul > > I see that Willi Brand has responded to your query with news that > Linde supply H2 with a dD value of -220 per mil. > > A few years ago Willi organised a 'ring test' (via Isogeochem) of > hydrogen gas samples supplied by Messer Griesheim. One of these > (ISO-TOP Ring-test Gas-A) has a dD value very close to SMOW, and we > have been using this (the cylinder contains about 1 litre at 12 bars) > ever since as our reference. > > Willi - is this gas still available? > > Tim Heaton > > NERC Isotope Geosciences Laboratory > British Geological Survey > Keyworth, Nottingham NG12 5GG, England > > Tel. +44(0)115 936 3401 > Fax +44(0)115 936 3302 > E-mail: [log in to unmask] Dear Tim et al, The gases for the ring test were a one time manufacture at Messer Griesheim. Although they still might have some of the flasks left from that filling procedure, I would not trust the numbers any more. (It is not wise anyway to rely on the isotope ratio value of a gas over years, in particular not for hydrogen). Under the brand name 'ISO-TOP' Messer Griesheim manufactures gases with an isotopic composition at or at least close to the desired delta value. Regards Willi P.S. Together with Ty Coplen I have summarized the results of the ring test for an IAEA Tec Doc report some time ago. I hope it will become available this year. WB -- ‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰ Willi A. Brand, Stable Isotope Laboratory [log in to unmask] Max-Planck-Institute for Biogeochemistry Tatzendpromenade 1a, 07745 Jena, Germany Tel +49-3641-643718 P.O.Box 100164, 07701 Jena, Germany Fax: +49-3641-643710 ‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰‰