Via offline reply, an easy fix: use the PAL loader software to "update" 
from backup.

As a reminder, always create a backup after creating a new method and 
also at least once a year. Can save a lot of time...

-Jim

On 10/20/2014 11:14 AM, Jim Palandri wrote:
> Hello All,
>
> While trying to change the GC PAL autosampler method moments ago, a new
> problem. Pressing the center button, and after scrolling down to the
> proper line and pressing the center button again, we get an electronic
> "chirp", and nothing more. Usually the entire line becomes highlighted,
> at which point we'd turn the jog dial to find the proper method, and
> press the the center button again. Restarting the system does not appear
> to help.
>
> Have any of you folks seen this behavior in the past?
>
> Thanks!
> -Jim
>

-- 