HI Andreas--

It's not the bands that are slipping.  The potentiometers are shot. Info 
is on the isogeochem wiki (I don't have the info with me).

take care,

gerry
On 1/8/2016 4:47 AM, Pack, Andreas wrote:
> Dear Isogeochemists,
>
> apart of having the annoying problem with the slipping rubber belt (that we are dealing with for the past years); since before christmas the “Hardware calibration” does not work anymore. When doing hardware calibration (iIrrespective of doing it in the Acquisition or Instrument control window); bellows are not moving and software freezes. We are using Isodat 3.
>
> The signal calibration still works well after adjusting the bellows potentiometer position manually.
>
> Did any of you had a similar problem (and a solution)?
>
> Andreas
>
> ————————————————————————————
> Prof. Dr. Andreas Pack