At 11:16 PM 9/3/99 +-200, you wrote: > Hello Flownetters, > >I would like to request some advice on the evaluation of carotid stenosis. >In our Hospital we measure stenosis by ultrasound comparing the diameter of the lumen with the diameter of the vessel at the point of maximum stenosis (real stenosis), but correlation with the results of arteriography is sometimes poor because they compare the diameter of the lumen with the distal diameter of the carotid. >Do you think that we must do the same by ultrasound? > >Carlos Nicolau > > Yes, Carlos. You should measure the percent stenosis by ultrasound the same way that your angiography department measures percent stenosis. By measuring the distal vessel diameter you are measuring the vessel at the more normal diameter. For example, if you measure the tightest area of stenosis in the proximal ICA, and this is where the carotid bulb is, you will get a falsely high percent stenosis if you also measure the diameter of the vessel here, because this is where the vessel is the widest. Measure the diameter of the vessel more distally, and you will get a more accurate percent stenosis by color angio. Deb Richert, RDMS, RVT La Crosse, WI